Repair and Service
Machine specific tools, software tools and auxiliary means
DD+DIS150.05E
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 3.2/48
Drystar 5302
Edition 1, Revision 0
Part 2: Resending the DICOM data
(1) Select 'Application - Comms
Recorder - Stop'
(2) Select 'Application - Comms
Recorder - Modify'
(3) In the Printer Snapshot edit
the desired parameters
(4) Select 'Application - Comms
Recorder - Replay': Now the
image is re-sent to the printer
and printed at the printer with
the adapted data.
Figure 49: Resending the DICOM
data