![AGFA Drystar 5302 Technical Documentation Manual Download Page 214](http://html.mh-extra.com/html/agfa/drystar-5302/drystar-5302_technical-documentation-manual_2866645214.webp)
DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 58 of 72
03-2011 Type
5366/100
Agfa Company Confidential
HC309020
: powersupplymodule.powerth
Type
Defect
Context
09
Action
Replace
Reason
PSU
module
defect
Cure
Replace
Part
MOD09
HC309022
: powersupplymodule.powermotor
Type
Defect
Context
09
Action
Replace
Reason1
Actuators power cables short circuited
Cure1
Check actuators power cables
Reason2
PSU
module
defect
Cure2
Replace
Part
MOD09
HC309024
: powersupplymodule.powerlogic
Type
Defect
Context
09
Action
Replace
Reason
PSU
module
defect
Cure
Replace
Part
MOD09
HC309025
: powersupplymodule.power3v3
Type
Defect
Context
09
Action
Replace
Reason
PSU
module
defect
Cure
Replace
Part
MOD09