DD+DIS150.05E
Controls, Connections and Set Up Procedures
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
Drystar 5302
Chapter 1 / 15
5 Film
Calibration
The purpose of this section is to show the tasks that are required for film
calibration
(1) Load the films the customer will
use in the upper and lower film
tray. Refer to note below.
(2) Select in key-operator menu
'6 Calibration - 1 Film
Calibration'
(3) Follow the instructions on the
display.
The calibration film will be
printed and measured
automatically by the printer
(see Figure 10)
(4) Repeat film calibration for
second input tray
536336as.cdr
Figure 10
The films delivered with the printer may only be used for test purposes.
2005-09-09 (Type