DRYSTAR 5302
Chapter 3.3 / 59 of 72
03-2011 Type
5366/100
Agfa Company Confidential
HC309026
: powersupplymodule.power12v
Type
Defect
Context
09
Action
Replace
Reason
PSU
module
defect
Cure
Replace
Part
MOD09
HC309027
: powersupplymodule.powern12v
Type
Defect
Context
09
Action
Replace
Reason
PSU
module
defect
Cure
Replace
Part
MOD09
MODULE : displaykeypadmodule
Elabel : Yes
HC310000
: displaykeypadmodule.elabel.present
Type
Failure
Context
W310000
Action
None
Reason1
Display Elabel not connected
Cure1
Connect Display elabel
Reason2
Display Elabel defect
Cure2
Replace
Part