![AGFA Drystar 5302 Technical Documentation Manual Download Page 212](http://html.mh-extra.com/html/agfa/drystar-5302/drystar-5302_technical-documentation-manual_2866645212.webp)
DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 56 of 72
03-2011 Type
5366/100
Agfa Company Confidential
MODULE : powersupplymodule
Elabel : Yes
HC309000
: powersupplymodule.temperature
Type
Defect
Context
E309000
Action
SeeManual
Reason1
PSU too hot
Cure1
Switch off printer and let it cool down
Reason2
PSU
module
defect
Cure2
Replace
Part
MOD09
HC309001
: powersupplymodule.elabel.present
Type
Failure
Context
W309001
Action
None
Reason1
PSU Elabel not connected
Cure1
Connect PSU elabel
Reason2
PSU module Elabel defect
Cure2
Replace
Part
MOD09
HC309002
: powersupplymodule.elabel.config
Type
Failure
Context
W309002
Action
None
Reason
PSU module Elabel defect
Cure
Replace
Part
MOD09