DD+DIS150.05E
Repair and Service
Machine specific tools, software tools and auxiliary means
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
Drystar 5302
Chapter 3.2/53
SW Installation
on empty HD
5366_Reg01_017_e.cdr
초
2
(1) Insert the CF card with the printer
"Emergency_Install" software in
the CF card reader
(2) Reboot the printer
(3) At question "Do you want to
preserve config?" select
"NO" in case you want to erase
the complete configuration
"YES" in case want to keep the
current configuration. Refer to
Figure 54 and to info below.
(4) Follow the further instructions on
the local keypad.
(5) Select menu "5 Restore
Configuration" to reload the
previous configuration when
finished.
Figure 55
The printer "Emergency_Install" software can be obtained via MEDNET,
GSO Library.
Caution:
•
Depending on the software patch version it might be possible, that
one of the boards gets re-programmed during the software upgrade.
•
In case this re-programming is interrupted by power loss, this specific
board may become unusable.
•
Do not reboot the printer before requested by a local keypad
message
When you choose during Emergency SW Installation to preserve the
configuration following steps are executed internally:
1. A backup of the printer configuration data is created and copied to
RAM
2. The software is copied from CF card to hard disk and inflated
3. The boards are programmed
4. The backup is restored from RAM
When you choose during Emergency SW Installation
not
to preserve the
configuration the configuration data are erased. The printer in this case is in
the same state like it has been shipped from factory
(default IP address 169.254.10.10)
2005-09-09 (Type
5366/100)
Agfa Company Confidential