DD+DIS150
Repair and Service
.05E
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Service - Calibration
Film calibration
1.
Internal densitometer calibration
2.
Thermal print head calibration
3.
Horizontal alignment
4.
Smart profile calibration
5.
Compliance film calibration
6.
Go back
Step Target values Measured values Step Target values Measured values
00
0.16
01
0.33
02
0.49
03
0.66
04
0.83
05
0.99
06
1.16
07
1.33
08
1.50
09
1.66
10
1.83
11
2.00
12
2.16
13
2.33
Service - Calibration
Compliance film calibration
14
2.50
15
2.66
Edition 1, Revision 1
DRYSTAR 5302
Chapter 3.6 / 30
05-2009 Type
5366/100
Agfa Company Confidential
16
2.83
17
3.00
Start calibration
Go back
Reset values
2.50
2.66
2.83
3.00
0.16
0.33
0.49
0.66
0.83
0.99
1.16
1.33
1.50
1.66
1.83
2.00
2.16
2.33
Service - Calibration
Compliance film calibration
Select densitometer:
Macbeth TR924
Gretag
X-Rite 301
X-Rite 310
Start calibration
Go back
Reset values
5366_reg36_02.CDR
Service - Calibration
Dmin:
0.20
Dmax:
3.15
Reflected Ambient Light
10
View Box Illumination
2000
Number of steps:
18 (EUREF)
Start calibration
Go back
Reset values
Service - Calibration
Compliance film calibration
1.
Upper tray
2.
Lower tray
Go back
Service - Calibration
Compliance film calibration
Actions which must have taken place with the same tray and with the same film pack:
1. First film calibration.
2. Profile calibration.
3. Second film calibration.
4. Remove film pack in other tray(s).
5. Printed DICOM-image.
6. Measurement of the densities.
Proceed
Go back
Service - Calibration
Compliance film calibration
1.
Reset compliance film calibration for upper tray (8x10)
2.
Reset compliance film calibration for lower tray (14x17)
3.
Proceed without resetting
Go back
Compliance film calibration
1
2
3
4
5
6
7
Compliance film calibration
Figure 18