DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 8 of 72
03-2011 Type
5366/100
Agfa Company Confidential
HC301007
: uppertraymodule.s1
Type
Failure
Context
W301007
Action
None
Reason1
S1 not connected
Cure1
Connect
S1
Reason2
Upper tray module defect
Cure2
Replace
Part
MOD01
HC301016
: upper.filmcal.required
Type
Failure
Context
W301016
Action
None
Reason
Upper tray needs calibration
Cure
Perform upper tray calibration
HC301018
: upperTag-unsupported
Type
Accidental
Context
Action
None
Reason
Wrong version in upper tray
Cure
Replace
filmpack
HC301019
: tag.history.insecure
Type
Defect
Context
E301019
Action
SeeManual
Reason
Tag history is corrupt
Cure
make Emergency sw install without preserving config