Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 42 of 72
03-2011 Type
5366/100
Agfa Company Confidential
MODULE : devilmodule
Elabel : Yes
HC308000
: devilmodule.dsh.connected
Type
Defect
Context
E308000
Action
SeeManual
Reason1
Connector P9 not connected
Cure1
Connect
P9
Reason2
MOD08 baseboard (Devil or Devius) module defect
Cure2
Replace
Part
MOD08
HC308001
: devilmodule.psu.connected
Type
Defect
Context
E308001
Action
SeeManual
Reason1
Connector P15 not connected
Cure1
Connect
P15
Reason2
MOD08 baseboard (Devil or Devius) module defect
Cure2
Replace
Part
MOD08
HC308002
: devilmodule.keypad.connected
Type
Defect
Context
E308002
Action
SeeManual
Reason1
Connector P8 not connected
Cure1
Connect
P8
Reason2
MOD08 baseboard (Devil or Devius) module defect
Cure2
Replace
Part
MOD08