DD+DIS148.05E
Installation Planning
1
2
3
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
7
Ambient conditions
7.1
During operation
Ambient temperature:
min.:
max.:
+15°C (59°F)
+30°C (86°F)
Relative humidity:
(non condensing)
min.:
max.:
20%
75%
7.2
During transport and storage
NOTE:
•
Climate conditions for storage must be in accordance with
EN60721-3-1-class 1K4
•
Climate conditions for transport must be in accordance with
EN60721-3-2(1993)-class 2K4
Ambient temperature
during storage:
min.:
max.:
- 25°C (-13°F)
+ 55°C (131°F)
Relative humidity
during storage:
min.:
max.:
10%
95%
Absolute humidity
during storage:
min.:
max.:
0.1 g/m³
35 g/m³
Rate of temperature
change during storage
max.: 1
°C/min.
Air pressure during
storage
Min.
max.:
70 kPa
106 kPa
Ambient temperature
during transport:
min.:
max.:
- 40°C (-40°F)
+ 70°C (158°F)
Relative humidity
during transport:
min.:
max.:
10%
95%
Edition 1, Revision 2
DRYSTAR 5302
Chapter 11 / 20
12-2007 5366/100
Agfa Company Confidential