![AGFA Drystar 5302 Technical Documentation Manual Download Page 225](http://html.mh-extra.com/html/agfa/drystar-5302/drystar-5302_technical-documentation-manual_2866645225.webp)
DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 69 of 72
03-2011 Type
5366/100
Agfa Company Confidential
MODULE : coolingfanmodule
Elabel : No
HC316000
: coolingfanmodule.thf1.connection
Type
Defect
Context
E316000
Action
SeeManual
Reason1
Coolingfan not connected
Cure1
Connect
coolingfan
Reason2
Coolingfan module defect
Cure2
Replace
Part
MOD12
HC316020
: coolingfanmodule.thf1.shortcircuit
Type
Defect
Context
E316020
Action
SeeManual
Reason1
Coolingfan not connected
Cure1
Connect
coolingfan
Reason2
Coolingfan module defect
Cure2
Replace
Part
MOD12