![AGFA Drystar 5302 Technical Documentation Manual Download Page 181](http://html.mh-extra.com/html/agfa/drystar-5302/drystar-5302_technical-documentation-manual_2866645181.webp)
DD+DIS150.05E
Repair and Service
Troubleshooting -
Error Codes
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.3 / 25 of 72
03-2011 Type
5366/100
Agfa Company Confidential
HC304023
: lowerpickupmodule.s16
Type
Failure
Context
W304023
Action
None
Reason1
Lower pickup home sensor S16 malfunctioning
Cure1
check lower pickup home sensor S16
Reason2
Lower pickup module defect
Cure2
Replace
Part
MOD04
HC304024
: lowerpickupmodule.s17
Type
Failure
Context
W304024
Action
None
Reason1
Lower pickup pickup sensor S17 malfunctioning
Cure1
check lower pickup pickup sensor S17
Reason2
Lower pickup module defect
Cure2
Replace
Part
MOD04
HC304025
: lowerpickupmodule.s19
Type
Failure
Context
W304025
Action
None
Reason1
Lower pickup passed sensor S19 malfunctioning
Cure1
check lower pickup passed sensor S19
Reason2
Lower pickup module defect
Cure2
Replace
Part
MOD04