DD+DIS150.05E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
2.5
Compliance Film Calibration as part of Afssaps / EUREF Calibration
Procedure
Purpose
Make film calibration this accurate, that it passes the acceptance criteria as defined
by Afssaps / EUREF:
The default film calibration procedure is not necessarily accurate enough.
When to be
performed
The compliance film calibration may be required as last step of the Afssaps / EUREF*
calibration procedure.
The Afssaps / EUREF compliance calibration procedure has to be performed in
following cases:
•
Before the Afssaps / EUREF acceptance test
•
Each 6 months, if the printer is under control of Afssaps / EUREF, and it did not
pass the Afssaps / EUREF calibration check (see section 2.5.6).
= Agence française de sécurité sanitaire des produits de santé
European Reference Organisation for Quality Assured Breast Screening and
Diagnostic
Services
Prerequiste
1) The compliance film calibration is only
available for DRYSTAR 5302
software
≥
2.0.0_C1.
2) A “TG18-PQC modified 4” image has
to be available on the modality.
The Test Pattern can be downloaded
from
Figure 10: TG18-PQC modified 4
Edition 1, Revision 1
DRYSTAR 5302
Chapter 3.6 / 17
05-2009 Type
5366/100