![AGFA Drystar 5302 Technical Documentation Manual Download Page 236](http://html.mh-extra.com/html/agfa/drystar-5302/drystar-5302_technical-documentation-manual_2866645236.webp)
Repair and Service
Electrical and mechanical codings, fuse tables and LED’s
DD+DIS150.05E
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 3.4 / 4
Drystar 5302
Edition 1, Revision 0
Agfa Company Confidential
(Type
5366/100)
2005-09-09
3.2
THEMIS Board (plugged into the DEVIL Board)
Name Led Color Function
D1
yellow
Software
controllable
V25 D2
green
2.5 for MPC107 host bridge and
FPGA
VCORE D3 green
CPU core voltage
CS1 D4
yellow
Chip select for portX bus
V33 D5
green
3.3V for most logic on board
FPGA D6
red
FPGA configuration done
If FPGA NOT configured => ALARM
D7
yellow
Software
controllable
VCC D8
green
5V for remaining logic on board
CS0 D9
yellow
Chip select for portX bus
D10
yellow
Software
controllable
CPU D11
yellow
CPU Bis Grant - indicates CPU
activity on the Bus