Repair and Service
Electrical and mechanical codings, fuse tables and LED’s
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Edition 1, Revision 0
Drystar 5302
Chapter 3.4 / 5
3.3 Power
Supply
536634ab.cdr
Figure 3
LED
Status
Color
Function
Overtemp
On
Off
Red
Temperature inside power supply
exceeded permissible limits. Operation
not possible*.
Temperature inside power supply
within permissible limit.
+ 3.3 V
On
Off
Green
+3.3 V present
+3.3 V not present
+ 5 V
On
Off
Green
+5 V present
+5 V not present
+12 V
On
Off
Green
+12 V present
+12 V not present
-12 V
On
Off
Green
-12 V present
-12 V not present
+24 V
On
Off
Green
+24 V present
+24 V not present
13..27 V
On
Off
Green
voltage between 13 V and 27 V
voltage < 13 V
*When the temperature inside the power supply exceeds the permissible limit,
printing can continue for up to 120 seconds. After this delay the power supply
switches off the power and checks the temperature every 20 seconds. As
soon as the temperature is below the permissible limit, it switches on again.