Maintenance
DD+DIS149.05E
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 9 / 14
DRYSTAR 5302
Edition 1, Revision 0
4.9 Electrical
check
The electrical check is split up in two steps:
Step
Task
Details see..
1
Perform the basic protective earth test
2
Perform the electrical check according to
National Regulations.
4.9.1
Basic Protective Earth Test
Agfa Company Confidential
(Type
5366/100)
2005-09-09
(1) Use a multimeter that is
capable to measure a
resistance < 1 Ohm
536631JN.cdr
2
(2) Put one probe at the ground
connector of the mains input
(3) Put the other probe at a few
metallic parts, the customer
could touch.
The electrical resistance has to
be
≤
0.2 Ohm.
Figure 14