7-2
IM 701210-05E
7.1 Starting/Stopping Waveform Acquisition
Conditions When Waveform Acquisition Cannot Be Started or Stopped
• Remote mode using the communication interface
• When printing or during auto setup
Setting the Waveform Acquisition Mode at Power ON
You can select whether to start waveform acquisition at power on. For example, select
ON if you wish to start waveform acquisition when power supply recovers after a power
failure. The initial setting is OFF.
ON:
Starts waveform acquisition at power on.
OFF:
Does not start waveform acquisition at power on.
START/STOP Key Response Time Setting
You can enter a setting to specify whether the instrument responds immediately when
the START/STOP key is pressed, or whether the instrument only responds if the key is
pressed for one second or longer.
Quick:
Immediate response
>1sec:
Responds when the key is pressed for one second or longer
Note
•
If you start waveform acquisition after changing the waveform acquisition conditions
indicated below, the data that had been stored in the acquisition memory up to that point is
cleared.
•
Setup conditions on the ACQ menu
•
Setup conditions of the module
•
Input coupling of the 701265 (TEMP/HPV) (when a change is made between TC
and some other setting) and the thermocouple type setting (when measuring
temperature)
•
Input coupling of the 701275 (ACCL/VOLT) (when a change is made between
ACCL and some other setting) and the sensitivity setting (when measuring
acceleration)
•
FV setting of the 701280 (FREQ) and offset setting
•
Trigger setup conditions
•
Setup conditions of the DUAL CAPTURE menu
•
T/div
•
A snapshot function that keeps the current displayed waveform on the screen is also
available. You can update the display without stopping the waveform acquisition (see the
next section).
•
For a description of the behavior when the Start Mode (Power On) is ON or OFF, see
appendix 9.
•
When waveforms are being acquired using roll mode display (see page 2-4), voice memo
can be recorded. For details, see section 7.9, “Using the Voice Memo Function.”