4-19
IM 701210-05E
Common Operations
4
4.5 Performing Auto Setup
Explanation
The auto setup function automatically sets the settings such as V/div, T/div, and trigger
level that are appropriate for the input signal.
Center Position after Auto Setup
0 V:
Sets the center to 0 V. This is useful when you wish to view the relative
relationship between the ground level and the waveform.
Offset:
Sets the center to the offset value. This is useful when you wish to view the
shape of the waveform by maximizing the amplitude.
Target Channel
Auto setup is performed on all channels or on the specified channel.
Applicable Modules
701250 (HS10M12), 701251 (HS1M16), 701255 (NONISO_10M12), 701260 (HV (with
RMS)), 701261 (UNIVERSAL), 701262 (UNIVERSAL (AAF)), and 701275 (ACCL/VOLT)
Canceling Auto Setup
Pressing the Undo soft key sets the DL750/DL750P back to the conditions that existed
immediately before auto setup. However, when you turn OFF the power switch, the
settings that existed immediately before auto setup are cleared. Therefore, the “Undo”
operation is not possible in this case. The settings that existed immediately before auto
setup are cleared if you execute initialization after auto setup.
Applicable Waveforms for Auto Setup
Frequency:
Approx. 50 Hz to 1 MHz
Absolute value of the input voltage:
Maximum value is between approximately 20 mV
(1:1) and ((maximum range)
×
10)
Type:
Repetitive waveform (that is not complex)
When the input coupling is set to DC
Note
The auto setup function may not work properly if the waveform includes DC components or
high-frequency components.