5-41
IM 701210-05E
Horizontal and Vertical Axes
5
As necessary, set the items of the preset (Logic 5V to EM Pickup) that you selected in
steps 2 and 3. If you select User (user-defined), set all the setup items arbitrarily. All
operations are done using the jog shuttle and SELECT.
• When changing the voltage range setting
Select the V Range (voltage range) from
±
1 V,
±
2 V,
±
5 V,
±
10 V,
±
20 V, and
±
50 V.
You can change the setting only when you set the preset to ZeroCross or User.
• When changing the input coupling setting
Set the coupling to DC or AC. You can change the setting only when you set the
preset to User.
• When changing the probe type setting
Select Probe and set the probe type to 1:1 or 10:1. You can change the setting only
when you set the preset to Logic 5V (3V/12V/24V), Pull-up 5V, ZeroCross, or User.
• When changing the bandwidth limit setting
Select the bandwidth limit from 100 Hz, 1 kHz, 10 kHz, 100 kHz, or Full. You can
change the setting on all presets. However, if the preset is set to AC100V or
AC200V, the bandwidth limit cannot be set to Full.
• When changing the threshold level setting
Set the threshold. You can change the setting only when you set the preset to User.
• When changing the hysteresis setting
Select Hys (hysteresis) from
±
1%,
±
2.5%, and
±
5%. You can change the setting on
all presets.
• When changing the slope setting
Set the slope to (rising) or (falling). You can change the setting only when you
set the preset to Logic 5V (3V/12V/24V), Pull-up 5V, or User.
• Setting the Chatter Elimination Function
Select Chatter Elimination and set the chatter elimination time in the range of 0 ms to
1000 ms. You can change the setting on all presets.
• Setting the pull-up
Set Pull Up to ON or OFF if you set the preset to Pull-up 5V.
The Pull Up setup menu appears only
when you set the preset to Pull-up 5V.
Voltage range
Input coupling
Probe type
Bandwidth limit
Threshold level
Hysteresis
Slope
Chatter elimination
function
5.19 Setting the Frequency (Number of Rotations, Period, Duty Cycle, Power Supply Frequency,
Pulse Width, Pulse Integration, and Velocity) Measurement