6-2
IM 701210-05E
Normal Mode
The display is updated only when the trigger conditions are met. The display is not
updated if the trigger does not occur. Therefore, to check the waveform or ground level
when no trigger is detected, use auto mode.
Single Mode
The display is updated once when the trigger conditions are met and the waveform
acquisition stops. In the time axis setting range in which the display mode is set to roll
mode, the display is set to roll mode. When a trigger occurs, the specified record length
of data is acquired and the displayed waveform stops.
Single (N) Mode
When the trigger condition is met, the waveform is acquired the specified number of
times and stops. This mode is used when acquiring a waveform using the sequential
store function.
Log Mode
In this mode, the trigger settings are disabled. The specified record length of data is
acquired once when acquisition is started, and the displayed waveforms are updated. If
the time axis setting is in the range in which the display mode is set to roll mode, the
display is set to roll mode.
Note
•
The trigger mode setting applies to both simple and enhanced triggers.
•
In the recorder mode of the DL750P (see chapter 9), select the trigger mode from Auto,
Repeat, Single, and Log. For details on the Repeat mode, see section 9.3.
6.1 Setting the Trigger Mode