6-35
IM 701210-05E
Triggering
6
6.15 Setting the Period Trigger (ENHANCED)
Setting the Bit Patterns of LOGIC A and LOGIC B
(Only When Using Logic Input Channels)
9.
Press the
Logic Pattern
soft key. The Logic Pattern setup dialog box appears.
10.
Use the
jog shuttle
and
SELECT
to select the pattern of each bit (Bit A1 to Bit
A8 and Bit B1 to Bit B8) from H, L, and X.
Setting the Trigger Level
11.
Press the
Select CH
soft key. The channel selection menu appears.
12.
Use the
jog shuttle
and
SELECT
to set the channel to assign the trigger level.
13.
Press the
Level
soft key.
14.
Use the
jog shuttle
and
SELECT
to set the trigger level.
15.
Likewise, set the trigger level for all channels for which patterns were assigned
in step 7.
Note
The trigger level setting applies to both simple and enhanced triggers.
Setting the Hysteresis
16.
Press the
Hysteresis
soft key to select
,
, or
.
Note
The pattern settings of condition B apply to all trigger types.