6-10
IM 701210-05E
When Measuring Temperature
Range
Resolution
K, E, J, T, L, U, N, R, S, B, W
0.1
°
C
Au7Fe
0.1 K
Setting the Trigger Slope
Select how the trigger source is to cross the specified level for activating the trigger from
the following three choices.
:
Activated when the trigger source changes from below the trigger level to above the
trigger level (rising).
:
Activated when the trigger source changes from above the trigger level to below the
trigger level (falling).
:
Activated on either a rising edge or falling edge.
Setting the Trigger Hysteresis
Sets a width to the trigger level so that triggers are not activated by small changes in the
trigger signal. Select the trigger hysteresis from the following three types.
• When Measuring Voltage
: Approx.
±
0.1 division of hysteresis around the trigger level.
: Approx.
±
0.5 division of hysteresis around the trigger level.
: Approx.
±
1 division of hysteresis around the trigger level.
• When Measuring Temperature
: Approx.
±
0.5
°
C (K,
°
F)
: Approx.
±
1
°
C (K,
°
F)
: Approx.
±
2
°
C (K,
°
F)
• When Measuring Strain
Hysteresis of approximately
±
2.5% of the range around the trigger level.
Hysteresis of approximately
±
12.5% of the range around the trigger level.
Hysteresis of approximately
±
25% of the range around the trigger level.
• When Measuring Acceleration
: Hysteresis of approximately
±
0.1 divisions of the range around the trigger level.
: Hysteresis of approximately
±
0.5 divisions of the range around the trigger level.
: Hysteresis of approximately
±
1 division of the range around the trigger level.
• When Measuring Frequency and Other Parameters on the Frequency Module
: Hysteresis of approximately
±
0.01 divisions of the range around the trigger
level.
: Hysteresis of approximately
±
0.5 divisions of the range around the trigger level.
: Hysteresis of approximately
±
1 division of the range around the trigger level.
Trigger is activated (When set to rising edge )
Hysteresis width
Trigger level
Setting the Hold Off
For details, see section 6.4, “Setting the Hold Off Time.”