3.
In the Assays step, select the assay or assays that you want to use in the run, and
then click Next.
Note:
·
After selecting an assay, the list is filtered to show compatible assays that can
be selected and run at the same time.
·
To create a new assay, see Chapter 3, “Create and manage assays (manager/
IMPORTANT!
Ensure that you select the assay that corresponds with the sample
type that you will use in the run. If you select a wrong assay when you create a
run plan, the instrument will use incorrect settings during the run, resulting in
invalid sequencing results. To determine which system-installed assay to use for
your sample type, see “System-installed assays“ on page 30.
Chapter 5
Create and manage run plans
Create a run plan for a sample run
5
58
Genexus
™
Integrated Sequencer User Guide