7.
In the Sample Plate step, review sample positions in the sample plate, modify
the concentration of samples, if needed, then click Next.
Click Bulk Edit Concentration to modify sample concentration of all samples
at one time, then click Update to return to the Sample Plate screen.
If a sample concentration is ≤1,024X of the target concentration for the assay,
which is displayed as a default value for each sample in the Sample Plate screen,
the sequencer automatically dilutes the sample to the target concentration during
the run. If a sample concentration is greater than this value, you must manually
dilute the sample to the target concentration, or to a value within range for
automated dilution before loading on the sample plate. For more information,
see “Dilute the samples (if needed) and load the sample plate—sample run“ on
Note:
The sample volume that is required for library preparation is not
adjustable. The volume depends on the number of primer pools in the assay,
sample type, and library chemistry. For more information, see “Dilute the
samples (if needed) and load the sample plate—sample run“ on page 69 for
specific sample volumes to load on the sample plate.
Chapter 5
Create and manage run plans
Create a run plan for a sample run
5
Genexus
™
Integrated Sequencer User Guide
61