IMPORTANT!
·
If you are using Ion AmpliSeq
™
library chemistry, leave the tube in position 2
empty and uncapped, but do not remove the tube from the carrier before
loading in the sequencer. Do not add a second Ion AmpliSeq
™
primer pool to
the position 2 tube.
·
If you are using Ion AmpliSeq
™
HD library chemistry, add the FWD and REV
primer pools to the appropriate tubes in the same carrier.
·
Ensure that no bubbles are introduced at the bottom of the tube when adding
the primer pool.
2.
If you do not install the primer pool tube carriers in the sequencer immediately,
cap the tubes that contain primer pools, then store the tube carriers on ice.
Remember to uncap all tubes before installing.
Load the sequencer and start a run
1.
Tap Run on the sequencer home screen to start the loading procedure.
Chapter 7
Load the sequencer and start a run
Load the sequencer and start a run
7
Genexus
™
Integrated Sequencer User Guide
75