Item
Description
When/How to use
Upload
Diagnostics
Allows upload of instrument diagnostics files for review by
support personnel.
Upload an instrument diagnostics file
after an aborted run, or if otherwise
directed to do so by Technical Support.
Note:
The file is pushed to Genexus
™
Software. To retrieve the file using
Genexus
™
Software, navigate to
(Settings)
4
About, click Instrument
Diagnostics, then follow the on-screen
instructions.
Shut Down
Access to "Shut Down" and "Reboot" commands.
If directed to do so by Technical
Support as part of a troubleshooting
procedure, or if the instrument will not
be used for more than three days.
Note:
It is not
necessary/recommended to shut down
the instrument overnight or over the
weekend. To power on the instrument
after a shut down, see “Power the
Appendix B
Touchscreen reference
Settings
B
128
Genexus
™