2.
In the Run Selection screen, select from the list the run plan that you want to
run.
Note:
If you select a run plan that requires more lanes than are available on a
currently installed chip, a dialog appears giving you the option to install a new
chip, or cancel. If you proceed with a new chip, a sequencing clean is performed,
then the sequencer prompts you to perform the Clear Deck, UV Clean, Load
Deck, Clear Sequencing Reagents, and Load Sequencing Reagents steps.
3.
In the Review Run screen, confirm the selected run plan, then tap Next.
The deck door opens automatically.
Note:
If the instrument vision system detects consumables loaded on the deck,
the sequencer prompts you to remove the consumables, then starts a UV Clean.
Chapter 7
Load the sequencer and start a run
Load the sequencer and start a run
7
76
Genexus
™
Integrated Sequencer User Guide