5.
In the QC step, enter Templating Control QC - CF1, Run QC, and Sample QC -
Library DNA/RNA parameters, then click Next. Leave parameters that are not
applicable to your assay as Not Set.
Parameters for QC pass/fail thresholds
Parameter
Description
Templating QC - CF-1
Average Reads Per
Lane
The average number of CF-1 reads per chip lane.
Base Call Accuracy The percentage of accuracy of the CF-1 calls aligned to the
CF-1 reference.
Mean AQ20 Read
Length
The mean length of CF-1 reads having ≥99% accuracy at each
position.
Run QC
Key Signal
The average signal after software processing for all ISPs that
identically match the library key (TCAG). A measure of the
efficiency of template amplification.
Percent Loading
The percentage of addressable wells on a chip lane that are
loaded with an ISP.
Raw Read Accuracy The percentage of raw reads mapping to the reference
sequence.
Chapter 3
Create and manage assays (manager/administrator)
Create a new assay (manager/administrator)
3
36
Genexus
™
Integrated Sequencer User Guide