Introduction, Features and Options
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
26
OPTION
OPTION
NUMBER
DESCRIPTION/NOTES
REFERENCE
SO
2
IZS Permeation Tubes
Replacement tubes for the IZS option; identical size/shape; different effusion rates.
Effusion Rate (@ 50°C)
Approximate
Concentation
Specified Flow Rate (of
indicated perm tube rate)
52C
796 ng/min
0.3-0.5 ppm
0.76 ± 5% lpm
N/A
52H
1592 ng/min
0.8 ppm
0.76 ± 50% lpm
N/A
52M
220 ng/min
150 ppb
0.56 ± 25% lpm
N/A
Each tube comes with a calibration certificate, traceable to a NIST
standard, specifying its actual effusion rate of that tube to within ± 5%
when immersed in a gas stream moving at the specified flow rate. This
calibration is performed at a tube temperature of 50°C.
Sections 3.5.4,
6.1.1.3 and 7.1.4
Communication Cables
For remote serial, network and Internet communication with the analyzer.
Type
Description
60A
RS-232
Shielded, straight-through DB-9F to DB-25M cable, about
1.8 m long. Used to interface with older computers or code
activated switches with DB-25 serial connectors.
Sections 3.4.8.3,
3.4.8.4, 5.1, 5.1.4,
5.1.5 and 5.2.2.7
60B
RS-232
Shielded, straight-through DB-9F to DB-9F cable of about
1.8 m length.
Sections 3.4.8.3,
and 3.4.8.4, 5.1,
5.1.4, 5.1.5 and
5.2.2.7
60C
Ethernet
Patch cable, 2 meters long, used for Internet and LAN
communications.
60D
USB
Cable for direct connection between instrument (rear panel
USB port) and personal computer.
Sections 3.4.8.2
and 5.1.2.1
Concentration Alarm Relay
Issues warning when gas concentration exceeds limits set by user.
61
F
our (4) “dry contact” relays on the rear panel of the instrument. This
relay option is different from and in addition to the “Contact Closures”
that come standard on all TML instruments.
RS-232 Multidrop
Enables communications between host compouter and up to eight analyzers.
62
Multidrop card seated on the analyzer’s CPU card.
Each instrument in the multidrop network requres this card and a
communications cable (Option 60B).
Section 3.4.8.4
Second Gas Sensors
Choice of one additional gas sensor.
65A
Oxygen (O
2
) Sensor
• Section 2.1 (specs)
• Section 3.5.5,
(pneumatic layout)
• Section 6.10.1
(calibration)
• Section 9.2 for
theory of operation
67A
Carbon Dioxide (CO
2
) Sensor
• Section 2.1 (specs)
• Section 3.5.6
(pneumatic layout)
• Section 6.10.2
(calibration)
• Section 9.3 (theory
of operation)
Special Options
Built in features, software activated
N/A
Maintenance Mode Switch
, located inside the instrument, places the
analyzer in maintenance mode where it can continue sampling, yet
N/A
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...