Teledyne ML - T100 UV Fluorescence SO2 Analyzer
APPENDIX A - Version Specific Software Documentation
05036 Rev D
A-25
APPENDIX A-4: Signal I/O Definitions
Table A-4: T100/M100E Signal I/O Definitions, Software Version G.4
SIGNAL NAME
BIT OR CHANNEL
NUMBER
DESCRIPTION
Internal inputs, U7, J108, pins 9–16 = bits 0–7, default I/O address 322 hex
0–7
Spare
AUX board digital outputs, default I
2
C address 30 hex
ELEC_TEST
3
0
1 = electrical test on
0 = off
OPTIC_TEST
3
1
1 = optic test on
0 = off
DARK_TEST
3
2
1 = dark test on
0 = off
PREAMP_RANGE_HI
3
3
1 = select high preamp range
0 = select low range
Internal outputs, U8, J108, pins 1
–
8 = bits 0
–
7, default I/O address 322 hex
ELEC_TEST
0
1 = electrical test on
0 = off
OPTIC_TEST
1
1 = optic test on
0 = off
PREAMP_RANGE_HI
2
1 = select high preamp range
0 = select low range
3–5
Spare
I2C_RESET
6
1 = reset I
2
C peripherals
0 = normal
I2C_DRV_RST
7
0 = hardware reset 8584 chip
1 = normal
Control inputs, U11, J1004, pins 1–6 = bits 0–5, default I/O address 321 hex
EXT_ZERO_CAL
0
0 = go into zero calibration
1 = exit zero calibration
EXT_SPAN_CAL
1
0 = go into span calibration
1 = exit span calibration
EXT_LOW_SPAN
2, 6
2
0 = go into low span calibration
1 = exit low span calibration
EXT_BKGND_CAL
4
3
0 = go into background calibration
1 = exit background calibration
4–5
Spare
6–7
Always
1
Control inputs, U14, J1006, pins 1
–
6 = bits 0
–
5, default I/O address 325 hex
0–5
Spare
6–7
Always
1
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...