
Getting Started
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
70
3.6.3. FUNCTIONAL CHECK
After the analyzer’s components have warmed up for at least 60 minutes, verify that the
software properly supports any hardware options that were installed.
For information on navigating through the analyzer’s software menus, refer to the menu
trees described in Appendix A.1.
Check to ensure that the analyzer is functioning within allowable operating parameters.
Appendix C includes a list of test functions viewable from the analyzer’s front panel as well
as their expected values.
These functions are also useful tools for diagnosing performance problems with your
analyzer (refer to Section 10.1.2).
The enclosed Final Test and Validation Data Sheet (P/N 04551) lists these values before the
instrument left the factory.
To view the current values of these parameters press the following control button
sequence on the analyzer’s front panel. Remember until the unit has completed its
warm up these parameters may not have stabilized.
1
This will match the currently selected units of measure for
the range being displayed.
2
Only appears if the CO
2
sensor option is installed.
3
Only appears if the O
2
sensor option is installed.
4
The
STABIL
function can be set to display data related to
any of the gasses the analyzer measures, e.g. (if either
the CO
2
option or the O
2
sensor option is installed).
5
Only appears if the IZS option is installed.
6
Only appears if analog output
A4
is actively reporting a
TEST FUNCTION
SAMPLE
RANGE=500.00 PPB SO2=XXXX
<TST
TST> CAL
SETUP
Toggle
<TST TST>
buttons
to scroll through list of
functions
RANGE=
[Value]
PPB
1
RANGE1=
[Value]
PPB
1
RANGE2=
[Value]
PPB
1
SO2 RNG=
[Value]
PPB
1, 2, 3
SO2 RN1=
[Value]
PPB
1, 2, 3
SO2 RN2=
[Value]
PPB
1, 2, 3
CO2 RNG=[Value]%
2
O2 RNG=[Value]%
3
STABIL=
[Value]
PPM
4
RSP=
[Value]
SEC
PRES=
[Value]
IN-HG-A
SAMP FL=
[Value]
CC/M
PMT =
[Value]
MV
NORM PMT=
[Value]
MV
UV LAMP=
[Value]
MV
LAMP RATIO=
[Value]
%
STR LGT=
[Value]
PPB
DRK PMT=
[Value]
MV
DRK LMP =
[Value]
MV
SLOPE=
[Value]
OFFSET=
[Value]
MV
CO2 SLOPE=[Value]
2
CO2 OFFSET=[Value]MV
2
O2 SLOPE=[Value]
3
O2 OFFSET=[Value]MV
3
HVPS =
[Value]
VOLTS
RCELL ON=
[Value]
SEC
RCELL TEMP=
[Value]
ºC
O2 CELL TEMP=[Value]ºC
2
BOX TEMP=
[Value]
ºC
PMT TEMP=
[Value]
º
IZS TEMP=
[Value]
ºC
5
PHT DRIVE =
[Value]
MV
TEST=
[Value]
MV
6
TIME=
[HH:MM:SS]
Figure 3-24:
Functional Check
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...