
Troubleshooting & Repair
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
282
06807A(DCNxxxx)
7. Check the voltages between the pairs of pins listed in Table 10-9. The result for each
pair should be equal and approximately 10% of the reading level recorded in Step 1.
Table 10-9: Example of HVPS Power Supply Outputs
If
HVPS
reading = 700 VDC
PIN PAIR
NOMINAL READING
KEY
5
6
7
8
9
10
11
1
2
3
4
1
2
70 VDC
2
3
70 VDC
3
4
70 VDC
4
5
70 VDC
5
6
70 VDC
6
7
70 VDC
7
8
70 VDC
8. Turn off the instrument power, and re-connect the PMT tube, and then re-assemble
the sensor.
If any faults are found in the test, the HVPS must be replaced. There are no user
serviceable parts inside the HVPS.
10.6.15. PNEUMATIC SENSOR ASSEMBLY
The pressure/flow sensor circuit board, located behind the sensor assembly, can be
checked with a voltmeter using the following procedure, which assumes that the wiring is
intact and that the motherboard and the power supplies are operating properly.
Measure the voltage across TP1 and TP2, it should be 10.0
0.25 V. If not, the board
may be faulty.
Measure the voltage across capacitor C2; it should be 5.0 ± 0.25 V. If not, the board
may be faulty.
10.6.16. SAMPLE PRESSURE
Measure the voltage across test points TP1 and TP4. With the sample pump disconnected
or turned off, this voltage should be 4500
250 mV. With the pump running, it should be
about 0.2 V less as the sample pressure drops by about 1 in-Hg-A from ambient pressure.
If this voltage is significantly different, the pressure transducer S2 or the board may be
faulty. A leak in the sample system to vacuum may also cause this voltage to be between
about 0.6 and 4.5. Ensure that the front panel reading of the sample pressure is at about 1
in-Hg-A less than ambient pressure.
10.6.17. IZS OPTION
The zero/span valves and IZS options need to be enabled in the software (contact the
factory on how to do this). Refer to Figure 3-19 and Figure 3-20 for a flow diagram with
zero/span valve or IZS option.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...