4.6.6. AUTO RANGE MODE (AUTO)
In AUTO range mode, the analyzer automatically switches the reporting range between
two user-defined ranges (low and high). The unit will switch from low range to high
range when either the SO
2
concentration exceeds 98% of the low range span. The unit
will return from high range back to low range once both the SO
2
concentration falls
below 75% of the low range span.
In
AUTO
Range mode the instrument reports the same data in the same range on both
the
A1
and
A2
outputs and automatically switches both outputs between ranges as
described above. Also, the
RANGE
test function displayed on the front panel will be
replaced by two separate functions:
RANGE1
: The
LOW
range setting for all analog outputs.
RANGE2
: The
HIGH
range setting for all analog outputs.
The high/low range status is also reported through the external, digital status bits (refer
to Section 5.3.1.1).
To set individual ranges press the following control button sequence.
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE:
SNGL
SNGL IND
AUTO
ENTR EXIT
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
EXIT
x 2 returns
to the main
SAMPLE
display
SETUP X.X
RANGE MODE:
AUTO
SNGL
IND AUTO
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
HIGH RANGE:
500.0 Conc
0 0 5 0 0 .0
ENTR
EXIT
SETUP X.X
LOW RANGE:
500.0 Conc
0 0 5 0 0 .0
ENTR
EXIT
Toggle the numeral
buttons to set the
LOW
and
HIGH
range value.
ENTR accepts the
new setting, EXIT
ignores the new
setting.
Figure 4-13:
SETUP RNGE
– Auto Range Mode
IMPORTANT
In AUTO range mode, the LOW and HIGH ranges have separate slopes and offsets for
computing SO
2
concentration. The two ranges must be independently calibrated.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...