Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Operating Instructions
81
Table 4-2: Test Functions Defined
DISPLAY
PARAMETER
UNITS
DESCRIPTION
RANGE
RANGE
- -
RANGE1
RANGE2
PPB, PPM,
UGM & MGM
The Full Scale limit at which the reporting range of the analyzer’s
ANALOG OUTPUTS is currently set.
THIS IS NOT the Physical Range of the instrument. Refer to Section
4.6 for more information.
If DUAL or AUTO Range modes have been selected, two RANGE
functions will appear, one for each range.
STABIL
STABILITY
mV
Standard deviation of SO
2
Concentration readings. Data points are
recorded every ten seconds. The calculation uses the last 25 data
points.
PRES
SAMPLE
PRESSURE
in-Hg-A
The current pressure of the sample gas as it enters the sample
chamber, measured between the SO
2
and Auto-Zero valves.
SAMP FL
SAMPLE FLOW
cm³/min
(cc/m)
The flow rate of the sample gas through the sample chamber. This
value is not measured but calculated from the sample pressure.
PMT
PMT Signal
mV
The raw output voltage of the PMT.
NORM PMT
NORMALIZED
PMT Signal
mV
The output voltage of the PMT after normalization for offset and
temperature/pressure compensation (if activated).
UV LAMP
Source UV Lamp
Intensity
mV
The output voltage of the UV reference detector.
LAMP
RATIO
UV Source lamp
ratio
%
The current output of the UV reference detector divided by the reading
stored in the CPU’s memory from the last time a UV Lamp calibration
was performed.
STR. LGT
Stray Light
ppb
The offset due to stray light recorded by the CPU during the last zero-
point calibration performed.
DRK PMT
Dark PMT
mV
The PMT output reading recorded the last time the UV source lamp
shutter was closed.
DRK LMP
Dark UV Source
Lamp
mV
The UV reference detector output reading recorded the last time the
UV source lamp shutter was closed.
SLOPE
SO
2
measurement
Slope
-
The sensitivity of the instrument as calculated during the last
calibration activity.
The slope parameter is used to set the span
calibration point of the analyzer.
OFFSET
SO
2
measurement
Offset
mV
The overall offset of the instrument as calculated during the last
calibration activity. The offset parameter is used to set the zero point
of the analyzer response.
HVPS
HVPS
V
The PMT high voltage power supply.
RCELL
TEMP
SAMPLE
CHAMBER TEMP
°C
The current temperature of the sample chamber.
BOX TEMP
BOX
TEMPERATURE
°C
The ambient temperature of the inside of the analyzer case.
PMT TEMP
PMT
TEMPERATURE
°C
The current temperature of the PMT.
IZS TEMP
1
IZS
TEMPERATURE
1
°C
The current temperature of the internal zero/span option. Only
appears when IZS option is enabled.
TEST
2
TEST SIGNAL
2
mV
Signal of a user-defined test function on output channel A4.
TIME
CLOCK TIME
hh:mm:ss
The current day time for DAS records and calibration events.
1
Only appears if Internal Gas Span Generator option is installed.
2
Only appears if analog output A3 is actively reporting a test function.
To view the TEST Functions press the following button sequence:
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...