Teledyne ML - T100 UV Fluorescence SO2 Analyzer
APPENDIX A - Version Specific Software Documentation
05036 Rev D
A-13
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE
DESCRIPTION
O2_FILT_ASIZE
10
Samples
10
1–500
O
2
moving average filter size in
adaptive mode.
O2_FILT_DELTA
10
% 2
0.1–100
Absolute change in O
2
concentration to
shorten filter.
O2_FILT_PCT
10
%
2 0.1–100 Relative change in O
2
concentration to
shorten filter.
O2_FILT_DELAY
10
Seconds
20
0–300
Delay before leaving O
2
adaptive filter
mode.
O2_DIL_FACTOR
10
— 1
0.1–1000
Dilution factor for O
2
. Used only if is
dilution enabled with
FACTORY_OPT
variable.
O2_CELL_SET
10
ºC
50 30–70
O
2
sensor cell temperature set point
and warning limits.
Warnings:
45–55
O2_STD_CELL_TEMP
10
ºK
323
1–500
Standard O
2
cell temperature for
temperature compensation.
O2_STD_CELL_PRESS
10
"Hg
28.50
1.00–50.00
Standard O
2
cell pressure for pressure
compensation.
USER_UNITS —
PPB,
PPM
3
PPB,
PPM,
UGM,
MGM,
PPM
3
,
MGM
3
Concentration units for user interface.
Enclose value in double quotes (“)
when setting from the RS-232
interface.
LAMP_DRIVE
6
mV
5000
0–5000
Lamp power setting.
LAMP_CAL
mV
3500
1000–5000
Last calibrated UV lamp reading.
LAMP_GAIN —
0.95
0.5–1.5
UV lamp compensation attenuation
factor.
BXTEMP_TPC_GAIN — 1,
0
12
0–10
Box temperature compensation
attenuation factor.
SPRESS_TPC_GAIN — 1
0–10
Sample pressure compensation
attenuation factor.
PMT_TARG_CONC
12
Conc 400
0.01–9999.99 Target SO
2
concentration during PMT
calibration.
PMT_UPDATE_PERIOD
12
Seconds 10
1–100
Period between HVPS gain updates
during PMT calibration.
PMT_CAL_TIMEOUT
12
Minutes
30
1–100
Maximum time for PMT calibration to
succeed.
HVPS_ADJUST
12
—
0
0–200
HVPS
gain adjustment.
HVPS_INTEG
12
Gain
5 0–500 Integral coefficient for adjusting HVPS
gain during PMT calibration.
HVPS_STABIL
12
—
1 0.1–10 HVPS gain must stabilize to within this
limit for PMT calibration to succeed.
PMT_ADJUST
12
—
0
0–65535
PMT gain adjustment.
SLOPE_CONST —
8,
6.25
3
0.1–10
Constant to make visible slope close to
1.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...