Teledyne ML - T100 UV Fluorescence SO2 Analyzer
APPENDIX A - Version Specific Software Documentation
05036 Rev D
A-15
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE
DESCRIPTION
PHYS_RANGE2 PPM
22,
5500
3
0.1–2500,
5–10000
3
High pre-amp range.
CONC_RANGE1 Conc
500,
5000
3
0.1–50000
D/A concentration range 1.
CONC_RANGE2 Conc
500,
5000
3
0.1–50000
D/A concentration range 2.
CO2_RANGE
11
%
15 0.1–500 CO
2
concentration range.
O2_RANGE
10
%
100
0.1–500
O
2
concentration range.
SAMP_FLOW_SET cc/m
700,
250
1+9
0–6000
Sample flow set point for flow
calculation and warning limits.
Warnings:
350–1200,
175–325
1+9
SAMP_FLOW_SLOPE —
1
0.5–1.5
Sample flow slope correction factor
(adjusted flow = measured flow x
slope).
VAC_SAMP_RATIO
3
— 0.53
0.1–2
Maximum vacuum pressure / sample
pressure ratio for valid sample flow
calculation.
SAMP_PRESS_SET "Hg 29.92 0–100
Sample pressure set point for pressure
compensation and warning limits.
Warnings:
15–35
VAC_PRESS_SET
3
"Hg 6
0–100
Vacuum pressure set point for
pressure compensation and warning
limits.
Warnings:
3–10
BOX_SET ºC
30
5–60
Box temperature warning limits. Set
point is not used.
Warnings:
8–50
PMT_SET ºC
7,
15
12
0–40
PMT temperature set point and
warning limits.
Warnings:
2–12,
2–20
12
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...