Getting Started
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
50
The software for this instrument is flexible enough to allow you to configure the alarms
so that you can have two alarm levels for each concentration.
SO
2
Alarm 1 = 20 PPM
SO
2
Alarm 2 = 100 PPM
SO
2
Alarm 1 = 20 PPM
SO
2
Alarm 2 = 100 PPM
In this example, SO
2
Alarm 1 and SO
2
Alarm 1 will both be associated with the “Alarm
2” relay on the rear panel. This allows you to have multiple alarm levels for individual
concentrations.
A more likely configuration for this would be to put one concentration on the “Alarm 1”
relay and the other concentration on the “Alarm 2” relay.
SO
2
Alarm 1 = 20 PPM
SO
2
Alarm 2 = Disabled
SO
2
Alarm 1 = Disabled
SO
2
Alarm 2 = 100 PPM
3.4.7.3.
“ALARM 4” RELAY
This relay is connected to the “range bit”. If the instrument is configured for “Auto
Range” and the reading goes up into the high range, it will turn this relay on.
3.4.8. CONNECTING THE COMMUNICATIONS INTERFACES
The T-Series analyzers are equipped with connectors for remote communications
interfaces:
Ethernet
(Section 3.4.8.1),
USB
(Section 3.4.8.2),
RS-232
(Section 3.4.8.3),
RS-232 Multidrop
(Section 3.4.8.4) and
RS-485
(Section 3.4.8.5). In addition to using
the appropriate cables, each type of communication method, except Ethernet, must be
configured using the SETUP>COMM menu, Section 5.1. Ethernet is DHCP-enabled by
default but can also be configured manually, which is the recommended method to
operatie with a static IP address.
3.4.8.1.
ETHERNET CONNECTION
For network or Internet communication with the analyzer, connect an Ethernet cable
from the analyzer’s rear panel Ethernet interface connector to an Ethernet port. Please
refer to Section 5.1.2 for a description of the default configuration and setup instructions.
Configuration
: Section 5.1.2
manual configuration: Section 5.1.2.1
automatic configuration (default): Section 5.1.2.2
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...