Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Getting Started
61
3.5.3. PNEUMATIC LAYOUT FOR ZERO/SPAN VALVES OPTION
Figure 3-19 shows the internal, pneumatic connections for a T100 with the zero/span
valve option installed.
FLOW
SENSOR
FLOW / PRESSURE
SENSOR PCA
SAMPLE
PRESSURE
SENSOR
V
AC
UU
M
M
AN
IF
O
L
D
FLOW
CONTROL
ASSY
EXHAUST TO OUTER
LAYER OF KICKER
SAMPLE/CAL
VALVE
SAMPLE GAS
INLET
ZERO AIR INLET
SPAN 1 INLET
ZERO/SPAN
VALVE
SAMPLE FILTER
Chassis
EXHAUST GAS
OUTLET
KICKER EXHAUST
TO PUMP
HYDROCARBON
SCRUBBER
(KICKER)
UV
LAMP
PMT
SAMPLE
CHAMBER
PUMP
Figure 3-19:
Pneumatic Layout with Zero/Span Valves Option
Table 3-10 describes the state of each valve during the analyzer’s various operational
modes.
Table 3-10: Zero/Span and Sample/Cal Valve Operating States
MODE
VALVE
CONDITION
SAMPLE
Sample/Cal
Open to SAMPLE inlet
Zero/Span
Open to ZERO AIR inlet
ZERO CAL
Sample/Cal
Open to zero/span inlet
Zero/Span
Open to ZERO AIR inlet
SPAN CAL
Sample/Cal
Open to zero/span inlet
Zero/Span
Open to SPAN GAS inlet
The state of the zero/span valves can also be controlled by any of the following means:
Manually from the analyzer’s front panel by using the
SIGNAL I/O
controls located under
the
DIAG
Menu (refer to Section 4.8.2),
By activating the instrument’s AutoCal feature (refer to Section 6.8),
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...