APPENDIX A - Version Specific Software Documentation
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
A-12
05036 Rev D
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE
DESCRIPTION
FILT_ASIZE Samples
20,
6
3
1–100
Moving average filter size in adaptive
mode.
FILT_DELTA PPM
0.02,
10
3
,
0.005
12
0.001–0.1,
1–100
3
Absolute change to trigger adaptive
filter.
FILT_PCT %
5,
10
2
,
1
12
1–100
Percent change to trigger adaptive
filter.
FILT_DELAY Seconds
180
0–300
Delay before leaving adaptive filter
mode.
FILT_ADAPT —
ON
OFF,
ON
ON enables adaptive filter; OFF
disables it.
NEG_CONC_SUPPRESS —
OFF
OFF,
ON
ON pegs negative concentrations at
zero; OFF permits negative
concentrations
DIL_FACTOR —
1
0.1–1000
Dilution factor if dilution enabled with
FACTORY_OPT
variable.
AGING_ENABLE
18
—
OFF
ON, OFF
ON enables aging offset and slope
compensation.
AGING_OFFSET_RATE
18
mV/day
0
-1.0–1.0
Aging offset rate of change per
day.
AGING_SLOPE_RATE
18
Change/
day
0
-0.01–0.01
Aging slope rate of change per
day.
CO2_DWELL
11
Seconds
1
0.1–30
Dwell time before taking each sample.
CO2_FILT_ADAPT
11
— ON
ON,
OFF
ON enables CO
2
adaptive filter; OFF
disables it.
CO2_FILT_SIZE
11
Samples
48
1–300
CO
2
moving average filter size in
normal mode.
CO2_FILT_ASIZE
11
Samples
12
1–300
CO
2
moving average filter size in
adaptive mode.
CO2_FILT_DELTA
11
% 2
0.1–10
Absolute change in CO
2
concentration
to shorten filter.
CO2_FILT_PCT
11
% 10 0.1–100 Relative change in CO
2
concentration
to shorten filter.
CO2_FILT_DELAY
11
Seconds
90
0–300
Delay before leaving CO
2
adaptive filter
mode.
CO2_DIL_FACTOR
11
—
1
0.1–1000
Dilution factor for CO
2
. Used only if is
dilution enabled with
FACTORY_OPT
variable.
CO2_STD_CELL_TEMP
11
ºK
323
1–500
Standard CO
2
cell temperature for
temperature compensation.
CO2_STD_CELL_PRESS
11
"Hg 28.50 1.00–50.00 Standard CO
2
cell pressure for
pressure compensation.
O2_DWELL
10
Seconds
1
0.1–30
Dwell time before taking each sample.
O2_FILT_ADAPT
10
— ON
ON,
OFF
ON enables O
2
adaptive filter; OFF
disables it.
O2_FILT_SIZE
10
Samples
60 1–500
O
2
moving average filter size in normal
mode.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...