Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Remote Operation & Advanced Features
135
5.1.6. COMM PORT TESTING
The serial ports can be tested for correct connection and output in the
COMM
menu.
This test sends a string of 256 ‘w’ characters to the selected COM port. While the test is
running, the red LED on the rear panel of the analyzer should flicker.
To initiate the test press, access the COMM Menu (refer to Figure 5-1), then press:
Select which
COMM port to
test.
SETUP X.X
COMMUNICATIONS MENU
ID INET
COM1
COM2
EXIT
SETUP X.X
COM1 : TEST PORT
<SET
TEST
EXIT
SETUP X.X
TRANSMITTING TO COM1
<SET
TEST
EXIT
SETUP X.X
COM1 MODE:0
SET>
EDIT
EXIT
SETUP X.X COM1 BAUD RATE:19200
<SET
SET>
EDIT
EXIT
EXIT
returns to
COMM
menu
Test runs
automatically
Figure 5-8: COMM
– COMM Port Baud Rate
5.2. USING THE DATA ACQUISITION SYSTEM (DAS)
The T100 analyzer contains a flexible and powerful, internal data acquisition system
(DAS) that enables the analyzer to store concentration and calibration data as well as a
host of diagnostic parameters. The DAS of the T100 can store up to about one million
data points, which can, depending on individual configurations, cover days, weeks or
months of valuable measurements. The data are stored in non-volatile memory and are
retained even when the instrument is powered off. Data are stored in plain text format
for easy retrieval and use in common data analysis programs (such as spreadsheet-type
programs).
The DAS is designed to be flexible, users have full control over the type, length and
reporting time of the data. The DAS permits users to access stored data through the
instrument’s front panel or its communication ports. Using APICOM, data can even be
retrieved automatically to a remote computer for further processing.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...