Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Getting Started
51
3.4.8.2.
USB CONNECTION
For direct communication between the analyzer and a PC, connect a USB cable between
the analyzer and desktop or laptop USB ports. (If this option is installed, the
COM2
port can only be used for Multidrop communication).The baud rate of the PC and the
analyzer must match (Section 5.1.5).
Configuration
: Section 5.1.3
3.4.8.3.
RS-232 CONNECTION
For
RS-232
communications with data terminal equipment (
DTE
) or with data
communication equipment (
DCE
) connect either a DB9-female-to-DB9-female cable
(Teledyne ML part number WR000077) or a DB9-female-to-DB25-male cable (Option
60A, Section 1.3), as applicable, from the analyzer’s rear panel RS-232 port to the
device. Adjust the DCE-DTE switch (Figure 3-4) to select DTE .or DCE as appropriate.
Configuration
: Sections 5.1, 5.1.4 and 5.1.5
IMPORTANT
Cables that appear to be compatible because of matching connectors may incorporate
internal wiring that makes the link inoperable. Check cables acquired from sources other
than Teledyne ML for pin assignments (Figure 3-12) before using.
RS-232
COM
P
ORT
C
ONNECTOR
P
IN
-
OUTS
Figure 3-12:
Rear Panel Connector Pin-Outs for RS-232 Mode
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...