Operating Instructions
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
120
4.8.7. LAMP CALIBRATION
An important factor in accurately determining SO
2
concentration is the amount of UV
light available to transform the SO
2
into SO
2
* (refer to Section 9.1.1). The T100
compensates for variations in the intensity of the available UV light by adjusting the SO
2
concentration calculation using a ratio (
LAMP RATIO
)that results from dividing the
current UV lamp (
UV LAMP)
intensity by a value stored in the CPU’s memory
(
LAMP_CAL
). Both LAMP Ration and UV Lamp are test functions viewable from the
instruments front panel.
To cause the analyzer to measure and record a value for
LAMP_CAL
, access the Signal
I/O from the DIAG Menu (refer to Figure 4-17), then press:
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Repeat Pressing
NEXT
until . .
.
DIAG
LAMP CALIBRATION
PREV NEXT
ENTR
EXIT
DIAG FCAL
LAMP CAL VALUE:4262.4 mV
4 2 6 2 .4 ENTR EXIT
E
NTR
accepts the
new value
EXIT
ignores the new
value
The value displayed is the
current output of the UV
source reference detector
Figure 4-33:
DIAG
– Lamp Calibration
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...