Teledyne ML - T100 UV Fluorescence SO2 Analyzer
APPENDIX A - Version Specific Software Documentation
05036 Rev D
A-11
APPENDIX A-2: Setup Variables For Serial I/O, Software Version G.4
Table A-1: Setup Variables, Software Version G.4
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE
DESCRIPTION
Low Access Level Setup Variables (818 password)
DAS_HOLD_OFF
Minutes
15
0.5–20
Duration of DAS hold off period.
TPC_ENABLE —
ON
OFF,
ON
ON enables temperature and pressure
compensation; OFF disables it.
RCELL_SET ºC
50
30–70
Reaction cell temperature set point and
warning limits.
Warnings:
45–55
IZS_SET
1
ºC
50
30–70
IZS temperature set point and warning
limits.
Warnings:
45–55
DYN_ZERO —
OFF
OFF,
ON
ON enables contact closure dynamic
zero; OFF disables it.
DYN_SPAN —
OFF
OFF,
ON
ON enables contact closure dynamic
span; OFF disables it.
CONC_PRECISION — 3
AUTO,
0,
1,
2,
3,
4
Number of digits to display to the right
of the decimal point for concentrations
on the display. Enclose value in double
quotes (“) when setting from the RS-
232 interface.
REM_CAL_DURATION
17
Minutes 20
1–120
Duration of automatic calibration
initiated from TAI protocol.
CLOCK_ADJ
Sec./Day
0
-60–60
Time-of-day clock speed adjustment.
Medium Access Level Setup Variables (929 password)
LANGUAGE_SELECT —
ENGL
ENGL,
SECD,
EXTN
Selects the language to use for the
user interface. Enclose value in double
quotes (“) when setting from the RS-
232 interface.
MAINT_TIMEOUT Hours
2
0.1–100 Time until automatically switching out
of software-controlled maintenance
mode.
CONV_TIME
—
33 MS
33 MS,
66 MS,
133 MS,
266 MS,
533 MS,
1 SEC,
2 SEC
Conversion time for PMT and UV
detector channels. Enclose value in
double quotes (“) when setting from the
RS-232 interface.
DWELL_TIME
Seconds
1
0.1–10
Dwell time before taking each sample.
FILT_SIZE Samples
240,
30
3
,
480
12
1–1000
Moving average filter size.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...