5.3.2.7.
COM PORT PASSWORD SECURITY
In order to provide security for remote access of the T100 , a LOGON feature can be
enabled to require a password before the instrument will accept commands. This is done
by turning on the
SECURITY MODE
(refer to Section 5.1.4). Once the
SECURITY
MODE
is enabled, the following items apply.
A password is required before the port will respond or pass on commands.
If the port is inactive for one hour, it will automatically logoff, which can also be achieved
with the LOGOFF command.
Three unsuccessful attempts to log on with an incorrect password will cause subsequent
logins to be disabled for 1 hour, even if the correct password is used.
If not logged on, the only active command is the '?' request for the help screen.
The following messages will be returned at logon:
LOGON SUCCESSFUL - Correct password given
LOGON FAILED - Password not given or incorrect
LOGOFF SUCCESSFUL - Connection terminated successfully
To log on to the T100 analyzer with
SECURITY MODE
feature enabled, type:
LOGON 940331
940331 is the default password. To change the default password, use the variable RS-
232_PASS issued as follows:
V RS-232_PASS=NNNNNN
Where N is any numeral between 0 and 9.
5.3.2.8.
APICOM REMOTE CONTROL PROGRAM
APICOM is an easy-to-use, yet powerful interface program that allows a user to access
and control any of Teledyne ML’s main line of ambient and stack-gas instruments from
a remote connection through direct cable, modem or Ethernet. Running APICOM, a user
can:
Establish a link from a remote location to the T100 through direct cable connection via RS-
232 modem or Ethernet.
View the instrument’s front panel and remotely access all functions that could be accessed
when standing in front of the instrument.
Remotely edit system parameters and set points.
Download, view, graph and save data for predictive diagnostics or data analysis.
Retrieve, view, edit, save and upload DAS configurations.
Check on system parameters for trouble-shooting and quality control.
APICOM is very helpful for initial setup, data analysis, maintenance and
troubleshooting. Figure 5-22 shows an example of APICOM being used to remotely
configuration the DAS feature. Figure 5-28 shows examples of APICOM’s main
interface, which emulates the look and functionality of the instruments actual front panel
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...