4.8.4.4.
ANALOG OUTPUT OFFSET ADJUSTMENT
Some analog signal recorders require that the zero signal to be significantly different
from the baseline of the recorder in order to record slightly negative readings from noise
around the zero point. This can be achieved in the T100 by defining a zero offset, a
small voltage (e.g., 10% of span), which can be added to the signal of individual output
channels by activating the
ANALOG I/O CONFIGURATION
MENU
from the DIAG
Menu (refer to Figure 4-17), then press:
DIAG AIO
RECORD OFFSET: 0 MV
+ 0 0 0 0
ENTR EXIT
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED:
NO
< SET
SET>
CAL
EXIT
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET>
EDIT
EXIT
Set the recorder
offset (in mV) of
the selected
channel
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 REC OFS: 0 mV
< SET
SET>
EDIT
EXIT
Pressing
ENTR
accepts the
new setting and returns to the
previous menu.
Pressing
EXIT
ignores the new
setting and returns to the
previous menu.
Press
SET>
to select the
analog output channel to
be configured. Then press
EDIT
to continue
DIAG AIO
CONC_OUT_2 OVERRANGE: ON
SET>
EDIT
EXIT
Figure 4-26:
Analog Output
– Offset Adjustment
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...