Getting Started
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
62
Remotely by using the external digital control inputs (refer to Section 5.3.1.2 and Section
6.7.1), or
Remotely through the RS-232/485 serial I/O ports (refer to Appendix A-6 for the
appropriate commands).
Sources of zero and span gas must be capable of supplying at least 1.55 L/min.
(maximum 2.5L/min). Both supply lines should be vented outside of the analyzer’s
enclosure. In order to prevent back-diffusion and pressure effects, these vent lines
should be between 2 and 10 meters in length.
3.5.4. PNEUMATIC LAYOUT FOR INTERNAL ZERO/SPAN (IZS) GAS
GENERATOR OPTION
Figure 3-20 shows the internal, pneumatic connections for the analyzer with the IZS
option installed.
EXHAUST TO OUTER LAYER
OF KICKER
V
AC
UU
M
M
AN
IF
O
L
D
CRITICAL
FLOW
ORIFICE
CRITICAL
FLOW
ORIFICE
Z
E
RO AIR
S
CR
UB
BE
R
SAMPLE/CAL
VALVE
Chassis
EXHAUST GAS
OUTLET
SAMPLE GAS
INLET
KICKER EXHAUST
TO PUMP
HYDROCARBON
SCRUBBER
(KICKER)
ZERO AIR INLET
UV
LAMP
PMT
SAMPLE
CHAMBER
FLOW
SENSOR
FLOW / PRESSURE
SENSOR PCA
SAMPLE
PRESSURE
SENSOR
PUMP
IZS
Permeation
Tube
SO
2
Source
ZERO/SPAN
VALVE
SA
M
PL
E
F
IL
T
ER
Figure 3-20:
Pneumatic Layout with IZS Options
The internal zero air and span gas generator (IZS) option includes a heated enclosure
(Section 3.5.4.1) for a permeation tube (permeation tube must be purchased separately;
see Section 1.3, in SO
2
IZS Permeation Tubes option), an external scrubber (Section
3.5.4.3) for producing zero air and a set of valves for switching between the sample gas
inlet and the output of the zero/span subsystem, functionally very similar to the valves
included in the zero/span valve option.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...