Remote Operation & Advanced Features
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
146
To configure the parameters for a specific data parameter, follow the instructions as
shown in Figure 5-14, then press:
SETUP X.X
0) PARAM=CONC1, MODE=AVG
PREV NEXT
INS DEL EDIT EXIT
SETUP X.X
PARAMETERS:CONC1
SET> EDIT
EXIT
SETUP X.X
SAMPLE MODE:AVG
<SET SET>
EDIT
EXIT
SETUP X.X
PARAMETERS: CONC1
PREV NEXT
ENTR
EXIT
Cycle through list of available
Parameters.
ENTR
accepts the new
setting and returns to the
previous menu.
EXIT
ignores the new setting
and returns to the previous
menu.
SETUP X.X
SAMPLE MODE: AVG
INST AVG SDEV MIN MAX
ENTR EXIT
Press the key for the desired mode
SETUP X.X
PRECISION: 1
<SET SET>
EDIT
EXIT
SETUP X.X
PRECISION: 1
1
ENTR EXIT
Set for 0-4
SETUP X.X
STORE NUM. SAMPLES: OFF
<SET
EDIT
EXIT
SETUP X.X
STORE NUM. SAMPLES: OFF
OFF
ENTR EXIT
Turn ON or OFF
<SET
Returns to
previous
Functions
Figure 5-15:
DAS
– Configuring Parameters for a Specific Data Parameter
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...