
Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Remote Operation & Advanced Features
145
5.2.2.4.
EDITING DAS PARAMETERS
Data channels can be edited individually from the front panel without affecting other
data channels. However, when editing a data channel, such as during adding, deleting or
editing parameters, all data for that particular channel will be lost, because the DAS can
store only data of one format (number of parameter columns etc.) for any given channel.
In addition, a DAS configuration can only be uploaded remotely as an entire set of
channels. Hence, remote update of the DAS will always delete all current channels and
stored data.
To modify, add or delete a parameter, follow the instruction shown in Figure 5-11, then
press:
Edit Data Parameter Menu
Edit Data Channel Menu
SETUP X.X
0)
CONC1: ATIMER, 2, 4032, R
PREV NEXT INS DEL
EDIT
PRNT
EXIT
Exits to the main
Data Acquisition
menu
SETUP X.X
NAME:CONC1
<SET
SET>
EDIT
EXIT
SETUP X.X
PARAMETERS : 2
<SET SET>
EDIT
EXIT
SETUP X.X
0) PARAM=CONC1, MODE=AVG
PREV NEXT
INS DEL EDIT EXIT
Press
SET>
key until…
SETUP X.X
EDIT PARAMS (DELETE DATA)
YES NO
NO returns to
the previous
menu and
retains all data.
Moves the
display between
available
Parameters
Inserts a new Parameter
before the currently
displayed Parameter
Deletes the Parameter
currently displayed.
Use to configure
the functions for
this Parameter.
Exits to the main
Data Acquisition
menu
YES will delete
all data in that
entire channel.
Figure 5-14:
DAS
– Editing DAS Parameters
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...