Teledyne ML - T100 UV Fluorescence SO2 Analyzer
Troubleshooting & Repair
06807A(DCNxxxx)
265
TEST
FUNCTION
NOMINAL VALUE(S)
POSSIBLE CAUSE(S)
LAMP RATIO
30 TO 120%
The current output of the UV reference detector divided by the reading stored in the
CPU’s memory from the last time a UV Lamp calibration was performed. Out of range
lamp ratio could be due to: malfunctioning UV lamp; UV lamp position out of alignment;
faulty lamp transformer; aging or faulty UV detector; dirty optical components; pin holes or
scratches in the UV optical filters; light leaks.
STR LGT
≤ 100 ppb / Zero Air
High stray light could be caused by: aging UV filter; contaminated reaction cell; light leak;
pneumatic leak.
DRK PMT
-50 to +200 mV
High dark PMT reading could be due to: light leak; shutter not closing completely; high
pmt temperature; high electronic offset.
DRK LMP
-50 to +200 mV
High dark UV detector could be caused by: light leak; shutter not closing completely; high
electronic offset.
HVPS
≈ 400 V to 900 V
Incorrect HVPS reading could be caused by; HVPS broken; preamp board circuit
problems.
RCELL TEMP
50ºC ± 1ºC
Incorrect temperature reading could be caused by: malfunctioning heater; relay board
communication (I
2
C bus); relay burnt out
BOX TEMP
Ambient
+
≈ 5ºC
Incorrect temperature reading could be caused by: Environment out of temperature
operating range; broken thermistor; runaway heater
PMT TEMP
7ºC ± 2ºC Constant
Incorrect temperature reading could be caused by: TEC cooling circuit broken; High
chassis temperature; 12V power supply
IZS TEMP
(OPTION)
50ºC ± 1ºC
Malfunctioning heater; relay board communication (I
2
C bus); relay burnt out
PRESS
Ambient
± 2 IN-HG-A
Incorrect sample gas pressure could be due to: pneumatic leak; malfunctioning valve;
malfunctioning pump; clogged flow orifices; sample inlet overpressure; faulty pressure
sensor
SLOPE
1.0 ± 0.3
Slope out of range could be due to: poor calibration quality; span gas concentration
incorrect; leaks; UV Lamp output decay.
OFFSET
< 250 mV
High offset could be due to: incorrect span gas concentration/contaminated zero air/leak;
low-level calibration off; light leak; aging UV filter; contaminated reaction cell; pneumatic
leak.
TIME OF DAY
Current Time
Incorrect Time could be caused by: Internal clock drifting; move across time zones;
daylight savings time?
10.1.3. USING THE DIAGNOSTIC SIGNAL I/O FUNCTIONS
The signal I/O parameters found under the diagnostics (DIAG) menu combined with a
thorough understanding of the instrument’s theory of operation (refer to Section 9) are
useful for troubleshooting in three ways:
The technician can view the raw, unprocessed signal level of the analyzer’s critical inputs and
outputs.
All of the components and functions that are normally under instrument control can be
manually changed.
Analog and digital output signals can be manually controlled.
This allows a user to systematically observe the effect of these functions on the operation
of the analyzer. Figure 10-2 shows an example of how to use the signal I/O menu to view
the raw voltage of an input signal or to control the state of an output voltage or control
signal. The specific parameter will vary depending on the situation. Please note that the
analyzer will freeze its concentration output while in the diagnostic signal I/O menu. This
is because manually changing I/O outputs can invalidate the instrument reading.
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...