T100 Instruction Manual
Warranty/Repair
Questionnaire T100
TELEDYNE ML CUSTOMER SERVICE
EMAIL:
PHONE: (303) 792-3300
TOLL FREE: (800) 846-6062 FAX: (303) 799-4853
068070000 Rev A
Pg: C-1
Customer: _______________________________________ Phone: _______________________________________
Contact Name: ___________________________________ Fax No: ______________________________________
Site Address: _____________________________________________________________________________________
Serial No.: _______________________________________
Firmware Revision: _____________________________
1. Are there any failure messages? ____________________________________________________________________
________________________________________________________________________________________________
Please complete the following table: (Note: Depending on options installed, not all test parameters shown below will be
available in your instrument)
* If option is installed
PARAMETER
RECORDED VALUE
UNITS
ACCEPTABLE VALUE
RANGE
PPB/PPM
50 PPB to 20 PPM Standard
STABIL
PPB
1 PPB with Zero Air
SAMP PRESS
In-Hg-A
Ambient ± 2 In-Hg-A
SAMPLE FLOW
cc/min
650 ± 10%
PMT SIGNAL WITH ZERO
AIR
mV
-20 to 150 mV
PMT SIGNAL AT SPAN
GAS CONC
mV
PPB/PPM
0-5000 mV
0-20000 PPB
NORM PMT AT SPAN
GAS CONC
mV
PPB/PPM
0-5000 mV
0-20000 PPB
UV LAMP
mV
1000 to 4800 mV
LAMP RATIO
%
30 to 120%
STR. LGT
PPB
≤
100 PPB/ Zero Air
DARK PMT
mV
-50 to 200 mV
DARK LAMP
mV
-50 to 200 mV
SLOPE
1.0 ± 0.3
OFFSET
mV
< 250 mV
HVPS
V
≈
400 – 900
RCELL TEMP
ºC 50ºC
±
1
BOX TEMP
ºC
A ~ 5
PMT TEMP
ºC
7ºC ± 2º Constant
IZS TEMP*
ºC 50ºC
±
1
ETEST
mV
2000 mV ± 1000
OTEST
mV
2000 mV ± 1000
Values are in the Signal I/O
REF_4096_MV
mV
4096mv±2mv and must be Stable
REF_GND
mV
0± 0.5 and must be Stable
2. What is the SAMPLE FLOW and SAMPLE PRESSURE with the sample inlet on rear of machine capped?
SAMPLE FLOW: __________________ cc/min SAMPLE
PRESSURE:
________________ In-Hg-A
Содержание T100
Страница 2: ......
Страница 4: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer iv This page intentionally left blank ...
Страница 6: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer vi This page intentionally left blank ...
Страница 10: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer x This page intentionally left blank ...
Страница 20: ...Teledyne ML T100 UV Fluorescence SO2 Analyzer xx ...
Страница 21: ...21 PART I GENERAL INFORMATION ...
Страница 77: ...77 PART II OPERATING INSTRUCTIONS ...
Страница 78: ...Getting Started Teledyne ML T100 UV Fluorescence SO2 Analyzer 78 ...
Страница 170: ...Remote Operation Advanced Features Teledyne ML T100 UV Fluorescence SO2 Analyzer 170 This page intentionally left blank ...
Страница 207: ...207 PART III TECHNICAL INFORMATION ...
Страница 208: ...208 06807A DCNxxxx ...
Страница 210: ...210 06807A DCNxxxx This page intentinally left blank ...
Страница 362: ...APPENDIX B T100 Spare Parts List T100 Instruction Manual B 2 068070000 Rev A ...
Страница 370: ...APPENDIX D ELECTRONIC SCHEMATICS T100 Instruction Manual D 2 068070000 Rev A ...
Страница 375: ...06807 Rev A D 5 TML 068070000 Rev A D 5 ...